AO2000 Series - Magnos206
AO2000 Series - Magnos27


AO2000 SERIES - MAGNOS206 (replaces MAGNOS106)

The Magnos206 is based on the paramagnetic behavior of oxygen and magneto-mechanical measuring principle.  Thanks to the short T90 time, the Magnos206 is also suitable for measuring rapid changes in the concentration of the sample gas.

Typical Applications

- Oxygen purity measurement
- Air separation plants
- Biogas monitoring
- Process gas monitoring
- Emission monitoring

Sample Components

- O2

Measuring Ranges

- Smallest measuring range: 0 - 0.5 Vol.% O2
- Largest measuring range: 0 - 100 Vol.% O2
- Measuring range suppression max. 1:100,  e.g. 99 - 100 Vol.% O2

Calibration

- Zero- and end-point calibration with nitrogen and air or test gas mixtures
- Single-point calibration with ambient air
- Automatic calibration via built-in pneumatic module or external valves

Measurement Principle

The measuring method of this analyzer module is based on the specific paramagnetic behaviour of oxygen.

Downloads

AO2000 Brochure  
 
AO2000 Data Sheet    
AO2000 Safety Brochure

 

AO2000 SERIES - MAGNOS27 (replaces MAGNOS17)

The Magnos27 is based on the paramagnetic behavior of oxygen and thermo-magnetic measuring principle.  The robust measuring cell means that the Magnos27 is especially resistant to vibrations and shocks.

Typical Applications

- Flue gas analysis
- Metal roasting plant off-gas analysis
- Cement flue gas analysis

Sample Components

- O2 in flue gas or nitrogen (N2)

Measuring Ranges

- Smallest measuring range: 0 - 3 Vol.% O2
- Largest measuring range: 0 - 100 Vol.% O2

Calibration

- Zero-point calibration with oxygen-free process gas or substitute gas
- End-point calibration with process gas having a known oxygen concentration or with
  substitute gas