PRODUCTS - AO2000 Series - MultiFID14, MultiFID14 NMHC

Back to Products


AO2000 Series - MultiFID14

AO2000 Series - MultiFID14 NMHC


AO2000 SERIES - MULTIFID14

The MultiFID14 is a flame ionization detector which measures the total content of organic carbon in the sample gas. For this purpose, organic substances are ionized in a hydrogen flame. The current of these ions is proportional to the organic carbon content. The analyzer is heated up to 200oC (392oF) and can be directly connected to a heated sample gas line; thus, no cold spots occur at any point. The MultiFID14 features self-monitoring, fault detection, logging and messaging functions. An automatic reset is also possible after fault correction.

Typical Applications

- Emission monitoring
- Process monitoring
- Measurement of volatile hydrocarbons (VOC) in water
- Purity measurements of O2, N2 and argon

Sample Components

- Hydrocarbons

Measuring Ranges

- Smallest measuring range: 0 - 5 mg org. C/m3
- Largest measuring range: 0 - 100,000 mg [100 g] org. C/m3
- Measuring range switching: manual; optional external control or automatic

Calibration

- Zero-point calibration with air or nitrogen
- Sensitivity calibration with propane or another hydrocarbon in air or nitrogen
- Automatic calibration via built-in zero gas and test gas valves

Downloads

AO2000 Brochure  
 
AO2000 Data Sheet    
AO2000 Safety Brochure

 

AO2000 SERIES - MULTIFID14 NMHC

The MultiFID14 NMHC (non-methane hydrocarbon) is a flame- ionization detector which measures the total contents of organic carbon with and without methane in the sample gas.  Via an internal solenoid valve, the sample gas is conducted through a catalyst in which all hydrocarbons are combusted into CO2 and H2O without methane.  Therefore, only the methane content is measured.  Then, the sample gas is supplied directly to the detector, leaving out the catalyst and the methane-free hydrocarbon concentration derived from the difference, can be displayed.

Typical Applications

- Emission monitoring
- Process monitoring
- Also for applications in which there is no O2 or H2O

Sample Components

- Hydrocarbons
- Methane
- Hydrocarbons without methane

Measuring Ranges

- Smallest measuring range:  0 - 5 mg org. C/m3
- Largest measuring range:    0 - 5,000 mg org. C/m3

Calibration

- Zero-point calibration with air or nitrogen
- Sensitivity calibration with propane or another hydrocarbon in air or nitrogen
- Automatic calibration via built-in zero gas and test gas valves
- Efficiency testing of catalyst with test gas C3H8 and CH4 in N2

Downloads

AO2000 Brochure  
 
AO2000 Data Sheet
AO2000 Safety Brochure

 