FEEDBACK LINKS SITE MAP CAREERS CEM Specialties Inc.Process & Emission Monitoring SpecialistsProducts | Services | Projects | Suppliers | What's New | About Company | Contact | Feedback | Links | Site Map | Careers |
|
|
Search For:
CEM SPECIALTIES INC.
1100 Dearness Drive
London, Ontario |
PRODUCTS
- EasyLine Series - EL3020-ZO23
ABB The trace oxygen analyzer ZO23 measures the gas concentration with a zirconium dioxide measuring cell. The measuring element consists of ceramics with a platinized surface, conducting oxygen ions at temperatures typically above 600oC. The measuring cell is catalytically inactive. The measuring method is especially advantageous to small measuring ranges of down to 1 ppm. This makes the analyzer particularly suitable for measuring oxygen in pure gases. Typical Applications
- Purity measurement Sample Components - O2 in nitrogen or argon Measuring Ranges
- Minimal
measuring range: 0...1 ppm Measurement Principle - Measurement of oxygen with catalytically inactive ZrO2 cell Dynamic Response - T90 time < 60 s when switching from sample to test gas Calibration - Manual or automatic calibration
Downloads |
Shortcuts
|
|
Copyright © 2006 CEM Specialties Inc. All rights
reserved. |
||